Formulation of objective indices to quantify machine failure risk analysis for interruptions in radiotherapy

Journal of Applied Clinical Medical Physics 22 巻 1 号 165-173 頁 2021-02-03 発行
アクセス数 : 271
ダウンロード数 : 63

今月のアクセス数 : 3
今月のダウンロード数 : 1
ファイル情報(添付)
JApplClinMedPhys_22_165.pdf 300 KB 種類 : 全文
タイトル ( eng )
Formulation of objective indices to quantify machine failure risk analysis for interruptions in radiotherapy
作成者
Nakano Hisashi
Ochi Yusuke
収録物名
Journal of Applied Clinical Medical Physics
22
1
開始ページ 165
終了ページ 173
抄録
Objectives: To evaluate the effect of interruption in radiotherapy due to machine failure in patients and medical institutions using machine failure risk analysis (MFRA).
Material and methods: The risk of machine failure during treatment is assigned to three scores (biological effect, B; occurrence, O; and cost of labor and repair parts, C) for each type of machine failure. The biological patient risk (BPR) and the economic institution risk (EIR) are calculated as the product of B and O (B×O) and C and O (C×O), respectively. The MFRA is performed in two linear accelerators (linacs).
Result: The multileaf collimator (MLC) fault has the highest BPR and second highest EIR. In particular, TrueBeam has a higher BPR and EIR for MLC failures. The total EIR in TrueBeam was significantly higher than that in Clinac iX. The minor interlock had the second highest BPR, whereas a smaller EIR. Meanwhile, the EIR for the LaserGuard fault was the highest, and that for the monitor chamber fault was the second highest. These machine failures occurred in TrueBeam. The BPR and EIR should be evaluated for each linac. Further, the sensitivity of the BPR, it decreased with higher T1=2 and α/β values. No relative difference is observed in the BPR for each machine failure when T1=2 and α/β were varied.
Conclusion: The risk faced by patients and institutions in machine failure may be reduced using MFRA.
Advances in knowledge: For clinical radiotherapy, interruption can occur from unscheduled downtime with machine failures. Interruption causes sublethal damage repair. The current study evaluated the effect of interruption in radiotherapy owing to machine failure on patients and medical institutions using a new method, that is, machine failure risk analysis.
著者キーワード
biological dose calculation
downtime
machine failure
言語
英語
資源タイプ 学術雑誌論文
出版者
American Association of Physicists in Medicine
Wiley
発行日 2021-02-03
権利情報
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of American Association of Physicists in Medicine.
出版タイプ Version of Record(出版社版。早期公開を含む)
アクセス権 オープンアクセス
収録物識別子
[ISSN] 1526-9914
[DOI] 10.1002/acm2.13126
[DOI] https://doi.org/10.1002/acm2.13126