Continuous Monitoring of Jugular Bulb Venous Oxygen Saturation for Evaluation of Cerebral Perfusion during Carotid Endarterectomy

アクセス数 : 964
ダウンロード数 : 152

今月のアクセス数 : 3
今月のダウンロード数 : 0
ファイル情報(添付)
HiroshimaJMedSci_47_133.pdf 493 KB 種類 : 全文
タイトル ( eng )
Continuous Monitoring of Jugular Bulb Venous Oxygen Saturation for Evaluation of Cerebral Perfusion during Carotid Endarterectomy
作成者
Nakagawa Itsuo
Shima Takeshi
Kawamoto Masaaki
Yuge Osafumi
収録物名
Hiroshima Journal of Medical Sciences
47
4
開始ページ 133
終了ページ 137
収録物識別子
[PISSN] 0018-2052
[EISSN] 2433-7668
[NCID] AA00664312
抄録
In this study we examined whether continuous monitoring of jugular bulb venous oxygen saturation (SjO2) is applicable for the evaluation of cerebral hypoperfusion during carotid endarterectomy (CEA). The subjects were 25 patients who underwent elective CEA under general anaesthesia. After the carotid stump pressure (SP) was measured, SjO2 and the somatosensory evoked potentials (SEP) were monitored during the carotid test clamping for 10 min.  

There was no alteration in cardiovascular and respiratory status during the test clamping. No correlation was observed between SEP amplitude and SP (r=0.16, p=0.25). However, at clamping, SjO2 decreased from 70 to 64% (p<0.01) with a reduction in SEP amplitude from 2.0 to 1.6μV (p<0.01). After declamping, SjO2 increased from 65 to 70% (p<0.01) with a recovery in SEP from 1.6 to l.9μV (p<0.01). The changes in SEP amplitude and SjO2 correlated (r=0.66, p<0.001).  

These results suggest that continuous monitoring of SjO2 is superior to SP measurement in the prediction of cerebral hypoperfusion caused by carotid clamping and applicable to CEA.
著者キーワード
Jugular bulb venous oxygen saturation
Cerebral ischaemia
Carotid clamp
Carotid endarterectomy
NDC分類
医学 [ 490 ]
言語
英語
資源タイプ 紀要論文
出版者
Hiroshima University Medical Press
発行日 1998-12
出版タイプ Version of Record(出版社版。早期公開を含む)
アクセス権 オープンアクセス
収録物識別子
[ISSN] 0018-2052
[NCID] AA00664312