Efficacy of the Pulse Pressure Generator during Cardiopulmonary Bypass Training Using the Extracorporeal Circulation Simulator

アクセス数 : 1164
ダウンロード数 : 172

今月のアクセス数 : 4
今月のダウンロード数 : 1
ファイル情報(添付)
HiroshimaJMedSci_67_71.pdf 4.69 MB 種類 : 全文
タイトル ( eng )
Efficacy of the Pulse Pressure Generator during Cardiopulmonary Bypass Training Using the Extracorporeal Circulation Simulator
作成者
Minami Shigeru
Ninomiya Shinji
収録物名
Hiroshima Journal of Medical Sciences
67
3
開始ページ 71
終了ページ 81
収録物識別子
[PISSN] 0018-2052
[EISSN] 2433-7668
[NCID] AA00664312
抄録
Objective: Cardiopulmonary bypass during cardiac surgery is an essential procedure, and the perfusionist needs to have sufficient education and training. Simulation training is suitable in such cases. We have developed a simulation system for cardiopulmonary bypass training (extracorporeal circulation simulator [ECCSIM]) and reported its efficacy. ECCSIM had no pulse pressure generator, so some perfusionists have mentioned that the operational feeling during training differs from that in real clinical cases.
In this study, we have developed a new pulse pressure generator and examined the efficacy of this system during cardiopulmonary bypass training using ECCSIM.
Materials and Methods: Results were observed as wave patterns during simulation of extracorporeal circulation with and without pulse pressure flow. Operational feeling during training of extracorporeal circulation was compared using a questionnaire survey, based on the Japanese version of the NASA Task Load Index (NASA-TLX), a subjective index, completed by seven perfusionists.
Results: With the addition of pulse pressure flow, fluctuation of arterial flow at low speed with the centrifugal pump increased, and operation of extracorporeal circulation became unstable. The questionnaire survey, including ‘Operational feeling of centrifugal pump’, ‘Feeling of afterload’, and ‘Display of pressure monitor’, showed results similar to that of the clinical situation using pulse pressure flow. The difficulty of simulator operation in extracorporeal circulation was significantly greater in the group with pulse pressure flow. Mental/physical load examined with NASA-TLX increased with pulse pressure flow.
Conclusion: Using a new pulse pressure generator with ECCSIM was effective in extracorporeal circulation training.
著者キーワード
cardiopulmonary bypass training
simulation
pulse pressure
NASA-TLX
NDC分類
医学 [ 490 ]
言語
英語
資源タイプ 紀要論文
出版者
Hiroshima University Medical Press
発行日 2018-09
権利情報
Copyright (c) 2018 Hiroshima University Medical Press
出版タイプ Version of Record(出版社版。早期公開を含む)
アクセス権 オープンアクセス
収録物識別子
[ISSN] 0018-2052
[ISSN] 2433-7668
[NCID] AA00664312
[DOI] 10.24811/hjms.67.3_71
[DOI] https://doi.org/10.24811/hjms.67.3_71