Noninvasive tracking of systolic arterial blood pressure using pulse transit time measured with ECG and carotid doppler signals with intermittent calibration
麻酔と蘇生 52 巻 1 号
25-28 頁
2016-03-20 発行
アクセス数 : 1007 件
ダウンロード数 : 135 件
今月のアクセス数 : 4 件
今月のダウンロード数 : 0 件
この文献の参照には次のURLをご利用ください : https://ir.lib.hiroshima-u.ac.jp/00039662
ファイル情報(添付) |
AnesthResus_52_25.pdf
1.9 MB
種類 :
全文
|
タイトル ( eng ) |
Noninvasive tracking of systolic arterial blood pressure using pulse transit time measured with ECG and carotid doppler signals with intermittent calibration
|
作成者 |
Fujita Yoshihisa
|
収録物名 |
麻酔と蘇生
Ansesthesia and Resuscitation
|
巻 | 52 |
号 | 1 |
開始ページ | 25 |
終了ページ | 28 |
抄録 |
We have developed a non-invasive blood pressure measurement system using pulse transit time (PTT) from the heart to the common carotid artery, measured by using an electrocardiogram (ECG) R-wave and carotid arterial Doppler signals at the anterior neck. In this study, we examined the validity of our system by comparing PTT derived systolic blood pressure (Dopp_SBP) with invasive radial systolic arterial pressure (Inv_SBP) with calibration every 15 min in the ICU setting.
Methods: 17 patients under invasive mechanical ventilation in the ICU were studied. Carotid arterial flow via an 8-MHz Doppler flow probe, ECG, and radial arterial pressure signals were transferred to a personal computer at a rate of 1 kHz and processed to calculate Dopp_SBP from PTT using our own calibration formula. Results: We recorded 3,770 pairs of Inv_SBP and Dopp_SBP in 17 patients. Inv_SBP ranged from 213 to 82 mmHg, and Dopp_SBP from 185 to 71 mmHg. The Bland-Altman plot of the comparison between Inv_SBP and Dopp_SBP revealed limits of agreement of ?20.1 to 17.7 mmHg (mean difference, -1.2 mmHg). There was a statistically significant close linear correlation between Inv_SBP and Dopp_SBP (y = 0.9494x + 7.5171, R2 = 0.8471, p < 0.0001). Conclusions: The results of the present study show that our system using Doppler ultrasound flow and ECG signals, is feasible for systolic blood pressure tracking over a longer interval if it is combined with intermittent calibration. |
著者キーワード |
doppler flow meter
blood pressure
monitor
|
内容記述 |
This work was supported by JSPS Kaken Grant Number 24592367.
|
NDC分類 |
医学 [ 490 ]
|
言語 |
英語
|
資源タイプ | 学術雑誌論文 |
出版者 |
広島麻酔医学会
|
発行日 | 2016-03-20 |
出版タイプ | Version of Record(出版社版。早期公開を含む) |
アクセス権 | オープンアクセス |
収録物識別子 |
[ISSN] 0385-1664
[NCID] AN00232251
|