Airway Scope(R) for Emergency lntubations : Usefulness of a New Video-laryngoscope
Hiroshima Journal of Medical Sciences 57 巻 3・4 号
99-104 頁
2008-12 発行
アクセス数 : 1367 件
ダウンロード数 : 169 件
今月のアクセス数 : 3 件
今月のダウンロード数 : 0 件
この文献の参照には次のURLをご利用ください : https://ir.lib.hiroshima-u.ac.jp/00034959
ファイル情報(添付) |
HiroshimaJMedSci_57_99.pdf
1.17 MB
種類 :
全文
|
タイトル ( eng ) |
Airway Scope(R) for Emergency lntubations : Usefulness of a New Video-laryngoscope
|
作成者 |
Ishida Makoto
Masuda Rieko
Tamura Tomoko
Takeda Taku
Tsumura Ryu
Shokawa Tomoki
Kondo Tomohiro
Sakai Hiroshi
Yamanoue Takao
|
収録物名 |
Hiroshima Journal of Medical Sciences
|
巻 | 57 |
号 | 3・4 |
開始ページ | 99 |
終了ページ | 104 |
収録物識別子 |
[PISSN] 0018-2052
[EISSN] 2433-7668
[NCID] AA00664312
|
抄録 |
The Airway Scope(R) (AWS-S100, Pentax, Tokyo, Japan) is a new video-laryngoscope, which has a specially designed blade and a portable, battery-powered, 2.4-inch built-in liquid crystal device (LCD) full-color video screen at the top of the handle. In this study, we tested the usefulness of the AWS for tracheal intubation in acute, urgent situations. Patients admitted to the Advanced Emergency and Critical Care Center at Hiroshima University Hospital and who required orotracheal intubations were prospectively enrolled. Tracheal intubations were performed using the AWS by residents (the novice group) or staff physicians (the experienced group), who received a brief instruction in the AWS. We enrolled 38 patients (23 males, 15 females; age, 60 ± 19 years). Intubations were attempted by 22 intubators (11 residents and 11 experienced personnel). The durations from inserting the blade via the oral cavity until observing the glottis (T1), inserting the tube into the trachea (T2), and confirming the chest rise (T3) were 22 ± 15, 34 ± 21and49 ± 27 sec, respectively. When the results were classified into t experienced and novice groups, T1, T2 and T3 were 17 ± 10 vs. 26 ± 17, 32 ± 23 vs. 36 ± 20, and 45 ± 25 vs. 53 ± 27 sec, respectively (the experienced vs. the novice group, n.s.). These results suggested that the AWS may be a suitable device particularly for less experienced personnel, such as novice Advanced Life Support providers.
|
著者キーワード |
Tracheal intubation
Video-laryngoscope
Emergency
|
NDC分類 |
医学 [ 490 ]
|
言語 |
英語
|
資源タイプ | 紀要論文 |
出版者 |
Hiroshima University Medical Press
|
発行日 | 2008-12 |
権利情報 |
(c) Hiroshima University Medical Press.
|
出版タイプ | Version of Record(出版社版。早期公開を含む) |
アクセス権 | オープンアクセス |
収録物識別子 |
[ISSN] 0018-2052
[NCID] AA00664312
|