タイトル ( eng )
The self-calibration of a retarding-type Mott spin polarimeter with a large collection angle
作成者
Iori Kazuyuki
Narita Hisashi
Sakamoto Kazuaki
Qiao Shan
収録物名
Review of Scientific Instruments
77
開始ページ 013101-1
終了ページ 013101-5
抄録
We have developed a compact retarding-type Mott spin polarimeter with a large collection angle at the Hiroshima Synchrotron Radiation Center and obtained a reliable value of the effective Sherman function by a self-calibration method. The spin polarization of secondary electrons was evaluated by extrapolation of the left-right scattering asymmetry at the zero energy-loss limit, which enabled us to obtain the effective Sherman function (Seff) and the figure of merit of our polarimeter at different operating conditions.
著者キーワード
electron spin polarisation
calibration
polarimeters
electron spectrometers
photoelectron spectroscopy