The self-calibration of a retarding-type Mott spin polarimeter with a large collection angle

Review of Scientific Instruments 77 巻 013101-1-013101-5 頁 2006-01-18 発行
アクセス数 : 925
ダウンロード数 : 230

今月のアクセス数 : 3
今月のダウンロード数 : 3
ファイル情報(添付)
RevSciInstrum_77_013101.pdf 253 KB 種類 : 全文
タイトル ( eng )
The self-calibration of a retarding-type Mott spin polarimeter with a large collection angle
作成者
Iori Kazuyuki
Narita Hisashi
Sakamoto Kazuaki
Qiao Shan
収録物名
Review of Scientific Instruments
77
開始ページ 013101-1
終了ページ 013101-5
抄録
We have developed a compact retarding-type Mott spin polarimeter with a large collection angle at the Hiroshima Synchrotron Radiation Center and obtained a reliable value of the effective Sherman function by a self-calibration method. The spin polarization of secondary electrons was evaluated by extrapolation of the left-right scattering asymmetry at the zero energy-loss limit, which enabled us to obtain the effective Sherman function (Seff) and the figure of merit of our polarimeter at different operating conditions.
著者キーワード
electron spin polarisation
calibration
polarimeters
electron spectrometers
photoelectron spectroscopy
言語
英語
資源タイプ 学術雑誌論文
出版者
American Institute of Physics
発行日 2006-01-18
権利情報
Copyright (c) 2006 American Institute of Physics.
出版タイプ Version of Record(出版社版。早期公開を含む)
アクセス権 オープンアクセス
収録物識別子
[ISSN] 0034-6748
[DOI] 10.1063/1.2162752
[NCID] AA00817730
[DOI] http://dx.doi.org/10.1103/PhysRevB.63.205108