低バックグラウンドGe検出器の遮蔽性能の改善

アクセス数 : 1341
ダウンロード数 : 712

今月のアクセス数 : 1
今月のダウンロード数 : 0
File
KJ00004291012.pdf 632 KB 種類 : fulltext
Title ( jpn )
低バックグラウンドGe検出器の遮蔽性能の改善
Title ( eng )
Background Reduction of Low Level Gamma-Ray Spectrometer
Creator
Shizuma Kiyoshi
Yokohata Kenji
Source Title
広島大学大学院工学研究科研究報告
Graduate School of Engineering, Hiroshima University
Volume 53
Issue 1
Start Page 87
End Page 93
Abstract
A new low-background shielding was installed for a Ge detector. Background gamma-ray spectrum was measured and compared with that of old shielding to investigate how improved the background counting rate. Additional improvements for the new shielding was installed to reduce background counting rate. It has been shown that shielding of the neck part of Ge detector cryostat and concrete floor are effective to reduce background contribution. The reduction factor of ^<40>K 1461 keV gamma-ray was much improved from 51 of old shield to 4290 of new shield.
Keywords
Ge spectrometer
Shielding
Low-level radioactivity
Reduction factor
Background gamma ray
NDC
Mechanical engineering [ 530 ]
Language
jpn
Resource Type departmental bulletin paper
Publisher
広島大学大学院工学研究科
国立情報学研究所
Date of Issued 2004-12
Publish Type Version of Record
Access Rights open access
Source Identifier
[ISSN] 1347-7218
[NCID] AA11700032