High-sensitive analysis of DNA fragments by capillary gel electrophoresis using transient isotachophoresis preconcentration and fluorescence detection
Journal of Chromatography A 1216 巻 17 号
3602-3605 頁
2009 発行
アクセス数 : 698 件
ダウンロード数 : 220 件
今月のアクセス数 : 4 件
今月のダウンロード数 : 4 件
この文献の参照には次のURLをご利用ください : https://ir.lib.hiroshima-u.ac.jp/00027705
ファイル情報(添付) |
JChromA_1216_3602.pdf
239 KB
種類 :
全文
|
タイトル ( eng ) |
High-sensitive analysis of DNA fragments by capillary gel electrophoresis using transient isotachophoresis preconcentration and fluorescence detection
|
作成者 |
Xu Zhongqi
Esumi Toshiaki
Ikuta Natsuki
|
収録物名 |
Journal of Chromatography A
|
巻 | 1216 |
号 | 17 |
開始ページ | 3602 |
終了ページ | 3605 |
抄録 |
In this report aimed on further development of a high-sensitivity capillary gel electrophoresis (CCE) method for analysis of DNA fragments, we firstly explored online transient isotachophoresis (tITP) preconcentration combined with fluorescence detection (FD). The fluorescence signal (excitation: 488 nm; emission: 590 nm) was generated using the intercalating dye of ethidium bromide (EB). It was found when the leading electrolyte (LE) was injected behind the sample zone, such a special tITP mode has significant advantages to solve the bubble formation issue and to improve the analytical performance stability. Two standard DNA samples, a 50 bp DNA step ladder and the phi X174/HaeIII digest. were used to evaluate the qualitative and quantitative abilities of the tITP-FD approach. A highly diluted sample (10,000-fold in the water, e.g. the phi X174/HaeIII digest diluted from 500 mu g/ml to the 50 ng/ml level) was enriched and detected; the LOD was down to 0.09 ng/ml for the 72 bp fragment, apparently improved more than 1000-fold in comparison with UV detection. Although the RSD of peak areas (n = 3) was around 15.5% for the sample was electrokinetically injected, good linearity of peak area response showed that the proposed method is suitable for quantitative analysis.
|
著者キーワード |
Capillary gel electrophoresis
DNA fragments
Transient isotachophoresis
Fluorescence detection
Ethidium bromide
|
NDC分類 |
化学工業 [ 570 ]
|
言語 |
英語
|
資源タイプ | 学術雑誌論文 |
出版者 |
Elsevier Science BV
|
発行日 | 2009 |
権利情報 |
Copyright (c) 2009 Elsevier B.V.
|