Continuous Monitoring of Jugular Bulb Venous Oxygen Saturation for Evaluation of Cerebral Perfusion during Carotid Endarterectomy
この文献の参照には次のURLをご利用ください : https://ir.lib.hiroshima-u.ac.jp/00037805
ID | 37805 |
本文ファイル | |
著者 |
Nakagawa, Itsuo
Shima, Takeshi
Kawamoto, Masaaki
Yuge, Osafumi
|
キーワード | Jugular bulb venous oxygen saturation
Cerebral ischaemia
Carotid clamp
Carotid endarterectomy
|
NDC |
医学
|
抄録(英) | In this study we examined whether continuous monitoring of jugular bulb venous oxygen saturation (SjO2) is applicable for the evaluation of cerebral hypoperfusion during carotid endarterectomy (CEA). The subjects were 25 patients who underwent elective CEA under general anaesthesia. After the carotid stump pressure (SP) was measured, SjO2 and the somatosensory evoked potentials (SEP) were monitored during the carotid test clamping for 10 min.
There was no alteration in cardiovascular and respiratory status during the test clamping. No correlation was observed between SEP amplitude and SP (r=0.16, p=0.25). However, at clamping, SjO2 decreased from 70 to 64% (p<0.01) with a reduction in SEP amplitude from 2.0 to 1.6μV (p<0.01). After declamping, SjO2 increased from 65 to 70% (p<0.01) with a recovery in SEP from 1.6 to l.9μV (p<0.01). The changes in SEP amplitude and SjO2 correlated (r=0.66, p<0.001). These results suggest that continuous monitoring of SjO2 is superior to SP measurement in the prediction of cerebral hypoperfusion caused by carotid clamping and applicable to CEA. |
掲載誌名 |
Hiroshima Journal of Medical Sciences
|
巻 | 47巻
|
号 | 4号
|
開始ページ | 133
|
終了ページ | 137
|
出版年月日 | 1998-12
|
出版者 | Hiroshima University Medical Press
|
ISSN | 0018-2052
|
NCID | |
言語 |
英語
|
NII資源タイプ |
紀要論文
|
広大資料タイプ |
学内刊行物(紀要等)
|
DCMIタイプ | text
|
フォーマット | application/pdf
|
著者版フラグ | publisher
|
部局名 |
医歯薬学総合研究科
|
他の一覧 |