Assessment of biological dosimetric margin for stereotactic body radiation therapy

Journal of Applied Clinical Medical Physics 21 巻 4 号 31-41 頁 2020-04-20 発行
アクセス数 : 272
ダウンロード数 : 59

今月のアクセス数 : 3
今月のダウンロード数 : 2
ファイル情報(添付)
JApplClinMedPhys_21_31.pdf 1.31 MB 種類 : 全文
タイトル ( eng )
Assessment of biological dosimetric margin for stereotactic body radiation therapy
作成者
Ozawa Shuichi
Shiinoki Takehiro
Tsubouchi Kento
収録物名
Journal of Applied Clinical Medical Physics
21
4
開始ページ 31
終了ページ 41
抄録
Purpose: To develop a novel biological dosimetric margin (BDM) and to create a biological conversion factor (BCF) that compensates for the difference between physical dosimetric margin (PDM) and BDM, which provides a novel scheme of a direct estimation of the BDM from the physical dose (PD) distribution.
Methods: The offset to isocenter was applied in 1‐mm steps along left‐right (LR), anterior‐posterior (AP), and cranio‐caudal (CC) directions for 10 treatment plans of lung stereotactic body radiation therapy (SBRT) with a prescribed dose of 48 Gy. These plans were recalculated to biological equivalent dose (BED) by the linearquadratic model for the dose per fraction (DPF) of d = 3–20 Gy/fr and α/β= 3 - 10. BDM and PDM were defined so that the region that satisfied that the dose covering 95% (or 98%) of the clinical target volume was greater than or equal to the 90% of the prescribed PD and BED, respectively. An empirical formula of the BCF was created as a function of the DPF.
Results: There was no significant difference between LR and AP directions for neither the PDM nor BDM. On the other hand, BDM and PDM in the CC direction were significantly larger than in the other directions. BCFs of D95% and D98% were derived for the transverse (LR and AP) and longitudinal (CC) directions.
Conclusions: A novel scheme to directly estimate the BDM using the BCF was developed. This technique is expected to enable the BED‐based SBRT treatment planning using PD‐based treatment planning systems.
著者キーワード
biological equivalent dose
dosimetric margin
LQ model
SBRT
言語
英語
資源タイプ 学術雑誌論文
出版者
American Association of Physicists in Medicine
Wiley
発行日 2020-04-20
権利情報
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
出版タイプ Version of Record(出版社版。早期公開を含む)
アクセス権 オープンアクセス
収録物識別子
[ISSN] 1526-9914
[DOI] 10.1002/acm2.12843
[PMID] 32141684