abstract
Abstract.The internal eye pressure is an important index for judging whether an eye suffers from glaucoma or not. The conventional eye pressure measurement is valid only under the condition that all subjects have the same structural eye stiffness. This paper challenges stiffness sensing of the human eye by pressing it with a contact probe. The displacement of the eye is captured by a camera with high resolution (see Fig.1). Experimental results show that the stiffness is roughly constant for each subject, while it differs by up to a factor of five between subjects. The method can detect subjects whose eye stiffness is smaller than average for further careful medical examination.
journal title
Proceedings - IEEE International Conference on Robotics and Automation
volume
Volume 2006
start page
2972
end page
2977
date of issued
2006
publisher
IEEE
issn
1050-4729
ncid
publisher doi
language
eng
nii type
Conference Paper
HU type
Conference Papers
DCMI type
text
format
application/pdf
text version
publisher
rights
Copyright (c) 2006 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted componet of this word in other works must be obtained from the IEEE.
relation url
department
Graduate School of Engineering