The self-calibration of a retarding-type Mott spin polarimeter with a large collection angle

Review of Scientific Instruments Volume 77 Page 013101-1-013101-5 published_at 2006-01-18
アクセス数 : 891
ダウンロード数 : 196

今月のアクセス数 : 0
今月のダウンロード数 : 3
File
RevSciInstrum_77_013101.pdf 253 KB 種類 : fulltext
Title ( eng )
The self-calibration of a retarding-type Mott spin polarimeter with a large collection angle
Creator
Iori Kazuyuki
Narita Hisashi
Sakamoto Kazuaki
Qiao Shan
Source Title
Review of Scientific Instruments
Volume 77
Start Page 013101-1
End Page 013101-5
Abstract
We have developed a compact retarding-type Mott spin polarimeter with a large collection angle at the Hiroshima Synchrotron Radiation Center and obtained a reliable value of the effective Sherman function by a self-calibration method. The spin polarization of secondary electrons was evaluated by extrapolation of the left-right scattering asymmetry at the zero energy-loss limit, which enabled us to obtain the effective Sherman function (Seff) and the figure of merit of our polarimeter at different operating conditions.
Keywords
electron spin polarisation
calibration
polarimeters
electron spectrometers
photoelectron spectroscopy