Newly Designed Eccentric Roller Type Total Artificial Heart
Hiroshima Journal of Medical Sciences 49 巻 1 号
7-14 頁
2000-03 発行
アクセス数 : 1101 件
ダウンロード数 : 144 件
今月のアクセス数 : 3 件
今月のダウンロード数 : 2 件
この文献の参照には次のURLをご利用ください : https://ir.lib.hiroshima-u.ac.jp/00037696
ファイル情報(添付) |
HiroshimaJMedSci_49_7.pdf
1020 KB
種類 :
全文
|
タイトル ( eng ) |
Newly Designed Eccentric Roller Type Total Artificial Heart
|
作成者 |
Wada Hideichi
Fukunaga Shintaro
Watari Masanobu
Orihashi Kazumasa
Matsuura Yuichiro
|
収録物名 |
Hiroshima Journal of Medical Sciences
|
巻 | 49 |
号 | 1 |
開始ページ | 7 |
終了ページ | 14 |
収録物識別子 |
[PISSN] 0018-2052
[EISSN] 2433-7668
[NCID] AA00664312
|
抄録 |
We have produced an eccentric roller type total artificial heart (ERTAH). As the first step in the development of this ERTAH, we conducted simulations such as a numerical simulation, a mock test, and an acute animal experiment using DeBakey roller pumps to analyze the left-right balance during its operation. The next step was redesigning the blood chambers to improve energy efficiency and implanting the ERTAH with an interatrial shunt into an animal for evaluation of the in vivo performance of the device. In the simulations, shunt flow through the bronchial arteries was approximately 500-600mlJmin, and the interatrial resistance was varied from 2.9-7.7 mmHg・min/liter. Redesigning the blood chambers resulted in the mock test in a 20% increase in energy efficiency, about a two-fold increase in cardiac output and improved durability compared to the previous type in the mock test. In the animal experiment the ERTAH operated with a left flow rate of 6.0 liters/min and a right flow rate of 5.4 liters/min. The interatrial shunt flow rate was 250-400m1Jmin. Creating an interatrial shunt is a useful method for balancing the blood flow between the left and right heart of the ERTAH. A decrease in friction resistance and the prevention of backward flow resulted in an increase in energy efficiency, cardiac output, and improved durability, in spite of downsizing the blood chambers.
|
著者キーワード |
Total artificial heart
Roller pump
Interatrial shunt
Left-right balance
|
内容記述 |
This work was supported in part by a Research Grant for Cardiovascular Disease 10C-2 from the Ministry of Health and Welfare, Japan.
|
NDC分類 |
医学 [ 490 ]
|
言語 |
英語
|
資源タイプ | 紀要論文 |
出版者 |
Hiroshima University Medical Press
|
発行日 | 2000-03 |
出版タイプ | Version of Record(出版社版。早期公開を含む) |
アクセス権 | オープンアクセス |
収録物識別子 |
[ISSN] 0018-2052
[NCID] AA00664312
|