A continuous-flow ATP amplification system for increasing the sensitivity of quantitative bioluminescence assay

Analytical Biochemistry 379 巻 1 号 116-120 頁 2008-08-01 発行
アクセス数 : 686
ダウンロード数 : 195

今月のアクセス数 : 2
今月のダウンロード数 : 1
ファイル情報(添付)
AnalBiochem_379_116.pdf 1.22 MB 種類 : 全文
タイトル ( eng )
A continuous-flow ATP amplification system for increasing the sensitivity of quantitative bioluminescence assay
作成者
Satoh Tetsuya
Shinoda Yasuharu
Alexandrov Maxym
収録物名
Analytical Biochemistry
379
1
開始ページ 116
終了ページ 120
抄録
We constructed a novel ATP amplification reactor using a continuous-flow system, and this allowed us to increase the sensitivity of quantitative bioluminescence assay by controlling the number of ATP amplification cycles. We previously developed a bioluminescence assay coupled with ATP amplification using a batch system. However, it was difficult to control the number of amplification cycles. In this study, ATP amplification was performed using a continuous-flow system, and significant linear correlations between amplified luminescence and initial ATP concentration were observed. When performing 4 cycles of continuous-flow ATP amplification, the gradient of amplification was 1.87N. Whereas lower quantifiable level was 500 pM without amplification, values as low as 50 pM ATP could be measured after amplification. Sensitivity thus increased 10-fold, with further improvements expected with additional amplification cycles. The continuous-flow system thus effectively increased the sensitivity of quantitative bioluminescence assay.
著者キーワード
Bioluminescence assay
ATP amplification
Continuous-flow system
NDC分類
生物科学・一般生物学 [ 460 ]
言語
英語
資源タイプ 学術雑誌論文
出版者
Academic Press Inc.
Elsevier Science
発行日 2008-08-01
権利情報
Copyright (c) 2008 Elsevier Inc.
出版タイプ Author’s Original(十分な品質であるとして、著者から正式な査読に提出される版)
アクセス権 オープンアクセス
収録物識別子
[ISSN] 0003-2697
[DOI] 10.1016/j.ab.2008.04.045
[NCID] AA00524867
[DOI] http://dx.doi.org/10.1016/j.ab.2008.04.045